Medical Gas Monitoring Systems

Area valve box

Technical Data:

Gases Types :

Standard measuring ranges:

Gas Consumption Counter by HGM

Basic features of the product:

Possible ranges:

Salient Features:

Oxygen flow and purity monitor

Oxygen flow and purity monitor in real-time. Device designed and developed to continuously monitor critical parameters of oxygen: pressure, oxygen purity.

This device is designed and developed to continuously monitor critical parameters of oxygen, such as: Pressure, Oxygen purity / concentration, and current flow / consumption.

This device is best utilized in oxygen concentrator / generator systems, where continuous monitoring or this crucial parameters is very important for medical facility.

Measuring ranges:

Inlet pressure: 0 – 16 bar

Basic parameters:

Dew point and CO monitor

To continuously monitor dew point, temperature and carbon monoxide level in compressed air.

Built in potential free relay and LOG module, that records events up to 10 days back.

Measuring ranges:

Relay (HF49F) max. 30 V, 3A DC; 48 V, 3A AC

Records last 250 entries, every 1 hour all alarms

Power supply: 230 V AC

MODBUS RTU (optional)

Area valve box

  • Designed to monitor pressure and flow rate of medical gases.
  • It can sum total usage of gas from the first start up, also has a history for last 6 months and every month is displayed separately.
  • Class IIb medical device.
Technical Data:
  • Working pressure: compressed gases 0 – 8 bar; vacuum -1 – 0 bar
  • Power supply: 12 V DC
  • Dimensions: 350x460x90 (WxHxD)
  • Connection to BMS and/or MODBUS RTU
Gases Types :
  • Oxygen
  • Nitrous Oxide
  • Carbon Dioxide
  • Compressed air
  • Nitrogen
  • Argon
  • Vacuum
Standard measuring ranges:
  • 0 – 100 l/min
  • 0 – 200 l/min
  • 0 – 500 l/min

Gas Consumption Counter by HGM

  • To measure the real time gas consumption of Gases in lpm and cumulative gas consumed in cubic meters
  • Ideal to measure the gas consumption by Individual Departments and Total Gas Consumption by entire Hospitals / Institutes
  • Accurate evaluation is the first step for greater productivity – We bring you easy to use , high efficiency devices to improve the output from your compressed air systems in a significant manner
  • This equipment is mainly designed to measure gas consumption at each ward
    Gas is measured proportionally and displayed in liters per minute
  • Precision of measurement is in range of +/- 5% within set measuring range. If consumption goes above set limit measurement precisionfalls down to +/- 12%
  • This unit also memorizes consumption for the past 6 months

Basic features of the product:

  • Color touch screen 4.3 “
  • Pressure measurement measured gas
  • Current flow
  • Aggregation of gas consumption
  • History of consumption for the last six months
  • Ball valves to cut off the gas supply
  • Accuracy of + / – 3%
Possible ranges:
  • 0 – 100 gallons per minute
  • 0 – 200 gallons per minute
  • 0 – 500 liters per minute
Salient Features:
  • Measuring Range from 0 – 500 lpm or as per request
  • Gas Consumption Counter is equipped with Pressure Monitor and with Alarm Function
  • Can also measure the consumption of Gases and retrieve data on Daily, weekly, monthly basis for 6 months from History log
  • Real-time Gas flow is displayed in LPM and cumulative gas consumption is displayed in cubic meters
  • Green LED indicator for System OK
  • Red LED indicator for Alarm condition
  • It is a Class IIb Medical Device
  • 12 V DC operated
  • WorkingPressure: 0 – 7 bar
  • Dimensions: (WxHxD) 350x460x90mm

Oxygen flow and purity monitor

Oxygen flow and purity monitor in real-time. Device designed and developed to continuously monitor critical parameters of oxygen: pressure, oxygen purity.

This device is designed and developed to continuously monitor critical parameters of oxygen, such as: Pressure, Oxygen purity / concentration, and current flow / consumption.

This device is best utilized in oxygen concentrator / generator systems, where continuous monitoring or this crucial parameters is very important for medical facility.

Measuring ranges:
  • pressure: 0 – 16 bar
  • concentration: 0 – 100%
  • flow: 0 – 200 l/min, 0 – 350 l/min, 0 – 500 l/min

Basic parameters:

  • Paramagnetic oxygen sensor
  • Thermal mass flow sensor

Inlet pressure: 0 – 16 bar

2 potential free relays

LOG – alarms history and gas consumption history

Dew point and CO monitor

To continuously monitor dew point, temperature and carbon monoxide level in compressed air.

Built in potential free relay and LOG module, that records events up to 10 days back.

Measuring ranges:

  • Dew point -100 – 100 ℃
  • CO 0 – 2000 ppm

Relay (HF49F) max. 30 V, 3A DC; 48 V, 3A AC

Records last 250 entries, every 1 hour all alarms

Power supply: 230 V AC

MODBUS RTU (optional)